Valleylab Force 1C Electrosurgical Unit
View Complete SpecificationsQuick Specs!
The Force 1C has a streamlined, compact design perfect for the requirements of outpatient surgery - as well as the power and performance features needed for a full range of general inpatient surgical procedures.
THREE BLENDED CUT MODES provide flexibility with varying degrees of hemostasis in the cut modes.
QUICK SET UP FEATURES reduce the need for readjustment between procedures. Microbipolar and monopolar modes are set independently, and the digital display allows easy confirmation of power levels.
ISOLATED OUTPUT DESIGN directs…
The Force 1C has a streamlined, compact design perfect for the requirements of outpatient surgery - as well as the power and performance features needed for a full range of general inpatient surgical procedures.
THREE BLENDED CUT MODES provide flexibility with varying degrees of hemostasis in the cut modes.
QUICK SET UP FEATURES reduce the need for readjustment between procedures. Microbipolar and monopolar modes are set independently, and the digital display allows easy confirmation of power levels.
ISOLATED OUTPUT DESIGN directs current return to the generator, reducing the risk of alternate site burns.
REM SAFETY is guaranteed with Valleylab's REM contact quality monitoring system, which continually monitors patient impedance levels. If a fault in the patient/return electrode contact is detected, the REM system automatically deactivates the generator - virtually eliminating the risk of burns under the return electrode.
Force 1C Generator
Technical Specifications
WEIGHT AND DIMENSIONS
OUTPUT WAVEFORMS
Output power changes by less than 5% as the line voltage varies from 85-135 volts
LOW FREQUENCY LEAKAGE (50/60 HZ)
Source current, patient leads, all outputs tied together
HIGH FREQUENCY LEAKAGE
Less than 150 mA rms
INPUT POWER REQUIREMENST
Operation range is 85 to 135 AC volts. Current is less than 5 ampers in cut and less than 3 amperes in coag.
POWER READOUTS
Agree with actual power into rated load to within ±20% or 5 watts, whichever is greater.
REM SYSTEM
Measurement Frequency: 140 kHz ±20 kHz
Measurement Current: 3 mA maximum
Acceptable Resistance Ranges:
COOLING
Convection, no fan
AUDIO VOLUME
The mode indicator tones are adjustable from 45 to 65 dBa
The alarm tones are not adjustable and are set to 65 dBa
Designed to meet UL and CSA specifications.