B. Braun McGaw Horizon NXT Infusion Pump
View Complete SpecificationsQuick Specs!
PRODUCT SPECIFICATIONS
Mechanism: Volumetric (non Drop-Counting), Positive Pressure Displacement Reservoir
Weight: Approximately 11.9 lb. (5.4 Kg)
Power Requirements: 90-132 VAC, 30 Watt, 47/63 Hz excluding accessories (domestic), 180 - 264 VAC, 30 Watts, 47/63 Hz excluding accessories (International)
Grounding Resistance : Meets UL Standard 544
Leakage Current: Meets or exceeds UL Standard 544, Section 27.2 and ANSI/AAMI Safe Current Limits (July 9, 1985)
Plug:Hospital Grade (3 pin)
Pole Size Range:0.75" to 1.25…
PRODUCT SPECIFICATIONS
Mechanism: Volumetric (non Drop-Counting), Positive Pressure Displacement Reservoir
Weight: Approximately 11.9 lb. (5.4 Kg)
Power Requirements: 90-132 VAC, 30 Watt, 47/63 Hz excluding accessories (domestic), 180 - 264 VAC, 30 Watts, 47/63 Hz excluding accessories (International)
Grounding Resistance : Meets UL Standard 544
Leakage Current: Meets or exceeds UL Standard 544, Section 27.2 and ANSI/AAMI Safe Current Limits (July 9, 1985)
Plug:Hospital Grade (3 pin)
Pole Size Range:0.75" to 1.25"
Battery:12 Volt sealed lead-acid battery for ambulatory use. Run time is dependent on the state of battery charge and the rate of fluid delivery.
Rate Range:Standard: (Primary and Piggyback) 0.1 to 999.9 ml/hr Micro: (Primary and Piggyback) 0.1 to 99.9 ml/hr
Volume to be Infused:0.1 to 9999.9 ml
Occlusion Pressure: 75, 100, 200, 300, 400, 500 and 750 mmHg Standard mode default is 300 mmHg.
Keep Open Rate:3 ml/hr for prescribed delivery rates of 3 to 999.9 ml/hr, or the continuation of the prescribed delivery rate for 0.1 to 2.9 ml/hr settings.
Volume Delivery Accuracy:+/- 5%
Fluid Types:All standard I.V. Fluids
Alarms:Air-In-Line, Container Empty, Door Open, Downstream Occlusion, Hold Time Exceeded, Low Battery, Low Flow From Container, Close Roller Clamp, System Error, Upstream Occlusion.
Air-In-Alarm: 0.09 cc
Hold Time Exceeded Alarm: Activated after three minutes on hold without user interaction (deactivated if the door of the pump is open)
Memory: Permanent data retention after the pump is turned off, unless the battery is depleted or disconnected, or unless the data Retention Defaults have been otherwise selected in the Biomed Options.
Service Directory
B. Braun McGaw Horizon NXT Infusion Pump