PRODUCT SPECIFICATIONS
Flow Rate Range:
- 0.1 to 999.9 ml/hr in 0.1 ml/hr increments (primary)
- 0.1 to 270.0 ml/hr in 0.1 ml/hr increments (secondary)
Volume To Be Infused Range:
- .1 to 9999.9 ml in 0.1 ml increments (primary)
- .1 to 999.9 ml in 0.1 ml increments (secondary)
Volume Infused Range:- 0.0 to 9999.9 ml in 0.1 ml increments
KVO Flow Rate:- 0.1-20.0 ml/hr in 0.1 ml/hr increments
Alarms:
- Air in Line
- Battery Depleted
- Channel Malfunction
- Computer Link Failure
- Flow Sensor Unplugged
- Hold Time Exceeded
- Instrument…
PRODUCT SPECIFICATIONS
Flow Rate Range:
- 0.1 to 999.9 ml/hr in 0.1 ml/hr increments (primary)
- 0.1 to 270.0 ml/hr in 0.1 ml/hr increments (secondary)
Volume To Be Infused Range:
- .1 to 9999.9 ml in 0.1 ml increments (primary)
- .1 to 999.9 ml in 0.1 ml increments (secondary)
Volume Infused Range:- 0.0 to 9999.9 ml in 0.1 ml increments
KVO Flow Rate:- 0.1-20.0 ml/hr in 0.1 ml/hr increments
Alarms:
- Air in Line
- Battery Depleted
- Channel Malfunction
- Computer Link Failure
- Flow Sensor Unplugged
- Hold Time Exceeded
- Instrument Malfunction
- Key stuck
- Latch Open
- No Upstream Flow Detected
- Occlusion Downstream
- Occlusion Upstream
- Primary Flow Detected during
Secondary:
- Set Out
- Set Up Time Exceeded
Dimensions
- Width 7.7 in/19.3 cm
- Height 8.6 in/21.8 cm
- Depth* 5.0 in/12.7 cm
- Weight 6.85 lbs./3.11 kg
- Power Cord 10 ft/3 m
Case: Impact resistant plastic
Power Requirements: 100- 120 V~,50/60 Hz, 0.5A,3-wire grounded system (40 watts Max)
Ground Current Leakage: Tested to UL Standard 544 and CSA C22.2 No. 125 for medical and dental equipment.
Battery:Rechargeable nickel cadmium. With a new, fully charged battery, the pump will operate for 4 hours nominal at 100 ml/hr for a two channel instrument operating on both channels simultaneously.
Environmental Conditions
Operating:
- Temperature Range 10ºC to 40ºC
(50ºF to 104ºF)
- Relative Humidity 15 to 95% Non-condensing
- Atmospheric Pressure 632 to 1031 mbar
Storage:
- Temperature Range -40ºC to 60ºC (-40ºF to 122ºF)
- Relative Humidity 5 to 95%
- Atmospheric Pressure 632 to 1031 mbar