Home . Alaris IVAC Signature Gold 7130 Infusion Pump

Alaris IVAC Signature Gold 7130 Infusion Pump

Item #: 7130
Available Refurbished
  • Description
  • Support Document
PRODUCT SPECIFICATIONS

Flow Rate Range:

- 0.1 to 999.9 ml/hr in 0.1 ml/hr increments (primary) - 0.1 to 270.0 ml/hr in 0.1 ml/hr increments (secondary)

Volume To Be Infused Range:

- .1 to 9999.9 ml in 0.1 ml increments (primary) - .1 to 999.9 ml in 0.1 ml increments (secondary)

Volume Infused Range:- 0.0 to 9999.9 ml in 0.1 ml increments

KVO Flow Rate:- 0.1-20.0 ml/hr in 0.1 ml/hr increments

Alarms:

- Air in Line - Battery Depleted - Channel Malfunction - Computer Link Failure - Flow Sensor Unplugged - Hold Time Exceeded - Instrument Malfunction - Key stuck - Latch Open - No Upstream Flow Detected - Occlusion Downstream - Occlusion Upstream - Primary Flow Detected during

Secondary:

- Set Out - Set Up Time Exceeded

Dimensions

- Width 7.7 in/19.3 cm - Height 8.6 in/21.8 cm - Depth* 5.0 in/12.7 cm - Weight 6.85 lbs./3.11 kg - Power Cord 10 ft/3 m

Case: Impact resistant plastic

Power Requirements: 100- 120 V~,50/60 Hz, 0.5A,3-wire grounded system (40 watts Max)

Ground Current Leakage: Tested to UL Standard 544 and CSA C22.2 No. 125 for medical and dental equipment.

Battery: Rechargeable nickel cadmium. With a new, fully charged battery, the pump will operate for 4 hours nominal at 100 ml/hr for a two channel instrument operating on both channels simultaneously.

Environmental Conditions

Operating:

- Temperature Range 10C to 40C (50F to 104F) - Relative Humidity 15 to 95% Non-condensing - Atmospheric Pressure 632 to 1031 mbar

Storage:

- Temperature Range -40C to 60C (-40F to 122F) - Relative Humidity 5 to 95% - Atmospheric Pressure 632 to 1031 mbar